KT 2000
The DYNEELAX is the most recent arthrometer released on the market (2020). It shares similar characteristics with the KT 1000, KT 2000 but is more precise while assessing knee laxity as it provides an objective evaluation of knee stability resulting from the automated anterior drawer test it performs. In addition, the DYNEELAX also offer rotation instability analysis.
Reproducibility of the tests is also an attribute of this device as several parameters incorporated in this tool enable medical practitioners to use the device in the pre / post surgery phases of the patient's medical management.
KT-1000 vs. GNRB
Comparison criteria’s |
KT-1000 | KT-2000 | GNRB | DYNEELAX |
Movement | Translation | Translation |
Translation |
Rotation + Translation |
Type of testing | Static: Displacement data at 134 N | Static: Displacement data at 134 N |
Dynamic: Force/Displacement curves (1 to 200N) |
Dynamic: Force/Displacement curves (1 to 200N) Torque / Rotation curves (1 to 8N) |
Tightening | Handmade: User dependent | Handmade: User dependent | Sensor: Data automatically collected | Sensor: Data automatically collected |
Device positioning | No data | No data | Ruler: Data on foot positioning collected | Ruler: Data on foot positioning collected |
Force applied | Handmade: User dependent | Handmade: User dependent | LDA software: Various controlled tibial forces can be chosen | LDA software: Various controlled tibial forces & torques can be chosen |
Displacement | Sound: Device beeps if displacement >3mm | Sound: Device beeps if displacement >3mm | Sensor: Data automatically collected | Sensor: Data automatically collected |
Results | Handmade: Hand written | Data saved on PC | LDA software: Data automatically saved | LDA software: Data automatically saved |
Test timing (min) | 2:15 | 2:03 | 2:58 | 10:00 |
Test Results | Anterior Knee Stability | Anterior Knee stability | Anterior Knee Stability |
Full knee analysis: Anterior Knee Stability Medial / Lateral Knee Stability |
Use | Pre-surgery | Pre-surgery | Pre-Surgery and Post-surgery |
Pre-Surgery and Post-surgery |
After carefully analyzing this tables chart, we can deduce that the post surgical use of the GNRB and DYNEELAX may represent the biggest asset of these arthrometers as it is possible to follow the state of the ACL graft.
This characteristic makes this device essential during ACL rehab as it practically guarantees the clinician that knee stability will be attained at some point since personnalized rehabilitation becomes possible for each unique patient.